How to fix airflow alarm
Airflow alarm errors displays, kindly provide a technical guide on how to fix it please.
@dako70477 if your isolette is a Air-Shields Isolette C400QT C450QT then go to this sie Air-Shields Isolette C400QT C450QT and download the service manual. "The airflow alarm is activated by either a fan failure or a short-circuited Air Flow probe" A sensor located below deck in the normal air path of the fan controls this alarm. If air flow stops due to a fan failure, the temperature of the self-heated sensor rises causing the Air Flow alarm Indicator on the front panel to light and produce a pulsating audible tone.
这个答案有帮助吗?
过去 24 小时: 0
过去 7 天: 3
过去 30 天: 6
总计 76
@dako70477 just to verify, your isolette is a Air-Shields Isolette C400QT C450QT?
由 oldturkey03 完成的