Repair guides and service information for the Puritan Bennett 840 model ventilator. It provides care for patients with respiratory failure or insufficiency. The Puritan Bennett 840 was originally produced in 1998, but has since undergone many critical software updates.

1 个问题 查看全部

Touch screen issue maybe GUI?

The bottom half of the touch screen lights up but has no graphics, it "beeps" when touched like it would normally, however you cannot see what you are touching. What is up with that?

回答此问题 我也有这个问题

这是一个好问题吗?

得分 1
添加一条评论

1个答案

最有帮助的答案

@tiffanykotzman either a loose cable on the display or a failed display. It may have lost contact with the LCD driver board which will then cause the error. Download the service manual from here Puritan Bennett 840 Service Ma.pdf Check around page 352 and on to see if you have a loose connection etc.

这个答案有帮助吗?

得分 1
添加一条评论

添加你的答案

Tiffany Kotzman 将永远感激不已
浏览统计数据: