跳转到主内容

Repair guides and service information for Medec Neptune anesthesia machines. The Medec Neptune Anesthesia is a ventilator designed for infants, adults and neonates (new-borns).

3 个问题 查看全部

Major leakage erro please help

Block Image

Hello guys

This error message appears and no tidal volume

Please help

回答此问题 我也有这个问题

这是一个好问题吗?

得分 0
添加一条评论

2个答案

最有帮助的答案

@zorro979 you most likely have issues witht he Tidal volume based on the leakage. I suggest you download the service manual from here and perform the Valve test which goes through the leakage checks. There are to many possibilities which require multiple tests and checkpoints.

这个答案有帮助吗?

得分 1
添加一条评论

Check for leakage of the endotracheal tube or LMA whatever you are using. If not the normal size, it will alarm leakage. What's the level of your oxygen? If more than 3L/Min , it will alarm leakage. Please set your PEEP to either 4 or 5.. do your machine check and correct the Fio2 21% and 100%.

这个答案有帮助吗?

得分 0
添加一条评论

添加你的答案

Waseem 将永远感激不已
浏览统计数据:

过去 24 小时: 0

过去 7 天: 1

过去 30 天: 12

总计 199