  1. Hamilton G5 Ventilation Unit Motherboard Replacement, Rail: 步骤 1 中的图像 1,1
    • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.

  2. Hamilton G5 Ventilation Unit Motherboard Replacement: 步骤 2 中的图像 1,1
    • Remove the Rail (B).

  3. Hamilton G5 Ventilation Unit Motherboard Replacement: 步骤 3 中的图像 1,1
    • The Standoff Brackets (C) remain with the Main Cover.

    • Place aside in a safe location.

  4. Hamilton G5 Ventilation Unit Motherboard Replacement, Main Cover: 步骤 5 中的图像 1,1
    • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.

    • Place aside in a safe location.

  5. Hamilton G5 Ventilation Unit Motherboard Replacement, Ventilation Unit Motherboard: 步骤 7 中的图像 1,1
    • First, we need to remove the ESM module. It is connected to 2 connector sockets. Disconnect and remove.

    • Take out the Ventilation Real-time Controller (VRC) SW chip

  6. Hamilton G5 Ventilation Unit Motherboard Replacement: 步骤 9 中的图像 1,1
    • Install the ESM Module, by connecting the connectors (A), and the VRC SW chip on the new board.

  7. Hamilton G5 Ventilation Unit Motherboard Replacement: 步骤 10 中的图像 1,1
    • Disconnect all cables and connections from the board to the unit.

  8. Hamilton G5 Ventilation Unit Motherboard Replacement: 步骤 11 中的图像 1,1
    • Remove the three Hex screws (B).

  9. Hamilton G5 Ventilation Unit Motherboard Replacement: 步骤 12 中的图像 1,1
    • Move the motherboard out then up away from the ventilation unit (C).

结论

To reassemble your device, follow these instructions in reverse order.

Dane Dormedy

于04/08/20注册

349 声望

创作了3篇指南

团队

Cal Poly, Team S7-G25, Paton Spring 2020 Cal Poly, Team S7-G25, Paton Spring 2020 的会员

CPSU-PATON-S20S7G25

6 名成员

创作了11篇指南

0条评论

添加评论

浏览统计数据:

过去 24 小时: 0

过去 7 天: 2

过去 30 天: 6

总计 254