SPECIFICATIONS

Selectable Loads and Ranges (Monopolar and Bipolar)

Load (Ohms)Range (Watts)
5050
100100
100300
200300
300300
400500
500500

Table Caption: This table lists various electrical loads in Ohms and their corresponding power ranges in Watts. It shows that for some load values, multiple power ranges are possible.

Isolation Test Range:

Current Range:

0 -50 Watts with 200-Ohm Load from Patient Plate, Active Electrode, and Chassis

0 -500 mA

Return Fault:

150 picofarads

Frequency Response : 0-5MHz

Frequency Response: (Filter Pushed) 0 - 50 kHz

Load Accuracy: + 2% at 50 ohms + 1.5% at 100 ohms + 1% 200 to 500 ohms

Meter Accuracy:

Power Output Accuracy:

  • 2%

  • 5% (Calibrated in Watts)

REM Test: Normal 24.6 Ohm Resistance, 137.0 Ohms with REM Switch Activated

Oscilloscope:

Power Requirements:

Measuring Technique:

Weight:

Isolated (Uncalibrated)

None

RF Thermocouples

5.4

kg

(12

Ibs)

Dimensions: 28 cm X 23 cm X 18 cm -(11"X 9.1" X7")

"