RETURN TO CD-ROM TABLE OF CONTENTS RETURN TO TECHNICAL SERVICE BULLETIN TABLE OF CONTENTS

DragerService

JULIAN TECHNICAL SERVICE BULLETIN #27

Product:

Julian

Date:

March 23, 2001

Re:

Corrective action - replacement of Man/Spont-Auto switchover diaphragm located in valve plate of the breathing system.

Reference Doc:

03/01 Maxeiner No.: 27 Released 3/9/2001

Reason/Rational:

To enhance the reliability of the Man/Spont switchover valve.

Solution/Procedure:

Replacement of individual diaphragms of diaphragm assembly for Man/Spont-Auto switchover.

Notice: When fitting the diaphragm(s), pay special attention to the correct fitting position (see drawing). Secure the screw with Loctite 221.

Devices affected:

All units manufactured prior to December 2000. (serial numbers prior to ARPN-xxxx)

Timing/When:

During next repair or Preventative Maintenance Certification visit, but action must be completed by March 2003 .

Cost:

Parts will be supplied free of charge up to the completion date of March 2003.

Time Required:

Without wait or travel times:

Fax: (215) 723-5935

DrägerService is a division of

Draeger Medical, Inc.

RETURN TO CD-ROM TABLE OF CONTENTS RETURN TO TECHNICAL SERVICE BULLETIN TABLE OF CONTENTS

Disassembly and assembly of the diaphragm assembly: 10 min Function test: 15 minutes

Total: 25 minutes

Ordering Info/Parts: M32514

  • (2x for each breathing system) Man/Spont diaphragm. Diaphragms can be ordered free-of-charge through Customer Service.

Checking/Testing:

Functional test: Carry out power on test, check ventilation parameters according to section 4.18- Ventilation Modes, of the Safety and Function test.

Distribution:

All Authorized Service Organizations, DragerService.

If you have any questions, please contact Technical Support by phone at 1-800-543-5047 or by fax at 1-215-721-5789.

DrägerService 3122 Commerce Drive Telford, PA 18969 Tel: (215) 721-5402 (800) 543-5047

Fax: (215) 723-5935

Fax: (215) 723-5935

Web: www.draegermedical.com

DrägerService is a division of

Draeger Medical, Inc.