Newport HT70 Pump Assembly Replacement

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。

你所需要的

  1. Newport HT70 Pump Assembly Replacement, Pump Assembly: 步骤 1 中的图像 1,1
    • Position the base assembly so that you may access it.

    • Using the #1 Phillips screwdriver, unscrew the four screws (SCR3219P in the photo) and the two screws (SCR3232P) securing the base plate.

  2. Newport HT70 Pump Assembly Replacement: 步骤 2 中的图像 1,1
    • The base plate will then come apart from the rest of the device with the pump assembly attached to it, as depicted in the following photo.

    • Inspect pump assembly for any damage or looseness.

  3. Newport HT70 Pump Assembly Replacement: 步骤 3 中的图像 1,1
    • If damage is found, replace the pump assembly by removing each tube individually and replacing as needed.

    • You may not need to replace the entire assembly if parts of it are still intact.

结论

To reassemble your device, follow these instructions in reverse order.