跳转到主内容

Hamilton-C3 Battery Replacement

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。

你所需要的

  1. Hamilton-C3 Battery Replacement, Battery: 步骤 1 中的图像 1,1
    • Connect the ventilator to an external power source. There is a two-prong cable that can be connected to a port on the backside of the ventilator.

    • It is vital that external power is present so that ventilation may continue even without a working battery in place.

  2. Hamilton-C3 Battery Replacement: 步骤 2 中的图像 1,2 Hamilton-C3 Battery Replacement: 步骤 2 中的图像 2,2
    • Open the battery door on the left side (facing the moniter) of the ventilator.

  3. Hamilton-C3 Battery Replacement: 步骤 3 中的图像 1,1
    • Twist the metal locking clip (indicated by red arrow) counterclockwise out of the locking position.

    • Use any flathead screwdriver to accomplish this task.

    • During reassembly, turn the metal locking clip (in red) clockwise, back into the locked position, to keep the battery secure.

    • Push the latch (component in red) upwards.

  4. Hamilton-C3 Battery Replacement: 步骤 4 中的图像 1,1
    • While pushing up the retaining clip, remove the battery (follow the blue arrow, ignore the white arrow).

    • During reassembly, push the new battery in until the latch (circled in red) locks down.

  5. Hamilton-C3 Battery Replacement: 步骤 5 中的图像 1,1
    • Turn the metal locking clip (in red) clockwise back into the locked position to keep the battery secure.

结论

To reassemble your device, follow these instructions in reverse order.