1. Disconnect the plug to the solar panel (yellow circle).
-
Then, grabbing the green connector and pull the battery connector from the solar charge controller (blue arrow).
-
2. Disconnect the plugs for the autosampler battery.
-
3. Disconnect the connector that plugs into the battery port of the sampler.
-
-
-
-
4. Make sure multimeter is measuring in DC current (red boxes).
-
Make sure multimeter is measuring in Volts (blue box).
-
5. Check the voltage of the lead acid autosampler battery. Fully charged lead acid battery should be around 13-14 V.
-
6. Check the voltage of the lithium ion node battery. Fully charged lithium ion battery should around 4 V.
-
-
-
7. First reconnect the autosampler battery (black box).
-
8. Then reconnect lithium ion battery in the node (blue box).
-
9. Secure solar charge controller to the side of the enclosure using velcro (red box).
-
Then plug in the solar panel last (orange box).
-