ventilator ltv 950 not working. power is on completed post test.

ventilator ltv 950 not working. power is on completed post test. but on vent check mode always. its not working normally. ventilator enters into vent check mode automatically on switching on.

回答此问题 我也有这个问题

这是一个好问题吗?

得分 0
添加一条评论

2个答案

Since this is a medical device, I think it will be hard to find information.

Based on your question, it is not getting feedback from the device that checks the vent and just hangs there. Similar to what Windows OS will do if it doesn't get feedback from a USB port.

Could be caused by a blockage. Any other cause is probably not user repairable.

这个答案有帮助吗?

得分 0
添加一条评论

you may have 2 situations, one can be select button is stuck and need to replace front panel, secondly check dip switch #5 to make sure it is off position (Switch toward outside ventilator). also #6&8 dip need to be off

这个答案有帮助吗?