VersaCare is a Hospital Bed manufactured by Hill-Rom. Model numbers P3200.

Why won't some of my P3200 beds not go as low as others?

I have a number of Versacare P3200 beds but some won't go all the way down while others will?

回答此问题 我也有这个问题

这是一个好问题吗?

得分 0
添加一条评论

1个答案

Be sure nothing is triggering the obstacle sensors. See if the blower assembly is not resting on the frame when in lower position. Could be Hilo motors need replacing

这个答案有帮助吗?

得分 0

1条评论:

Sometimes the tan aux outlet cable causes the obstacle sensors to trigger.

I have seen the bowers assembly be in a position such it will rest on the frame. You can sometimes see chips in the paint. This will also cause the scale to not work if attempting to weigh the patient when the bed is lowered down.