  1. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement, PIM Board Kit: 步骤 1 中的图像 1,1
    • Loosen but do not remove the (4) 8-32 X 3 screws on the outer air intake

  2. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 2 中的图像 1,2 Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 2 中的图像 2,2
    • Loosen and remove the (2) 10-32 Keps nuts and the (4) 6-32 X 2 screws. Remove the back case by lifting from the ventilator.

  3. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 3 中的图像 1,1
    • Remove the (2) 4-40 X ¼ screws on the Dovetail Mounting Bracket and disconnect the ribbon cable on the PIM PCB by simultaneously applying pressure on the two locking “ears”.

  4. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 4 中的图像 1,1
    • Flip the ventilator over and remove the front case assembly by lifting it straight up away from the ventilator module.

  5. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 5 中的图像 1,1
    • Remove the battery compartment cover by unscrewing the (4) 6-32 X 5/16 screws.

    • Remove the battery by unscrewing the (4) 6-32 X 2 ¼ screws and detaching the plug from its locking latch.

  6. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 6 中的图像 1,1
    • Disconnect the Power Input cable from the PIM PCB by pressing on the locking latch and pulling the cable straight up from the connector.

  7. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 7 中的图像 1,2 Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 7 中的图像 2,2
    • Loosen and remove the(5) 4-40 X 5/16 screws holding the PIM PCB to the ventilator module. Position SPO2 Insulator out of the way (Do Not Fold) and insert screwdriver through the holes to aid in removing screws.

  8. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 8 中的图像 1,1
    • Lift the defective PIM Board out of the ventilator module.

  9. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 9 中的图像 1,3 Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 9 中的图像 2,3 Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 9 中的图像 3,3
    • Place the new PIM Board Unto the vent module (make sure male header pins are inserted correctly into the mating header.

  10. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 10 中的图像 1,2 Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 10 中的图像 2,2
    • Secure the PIM Board with the (5) 4-40 x 5/16 screws.

  11. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 11 中的图像 1,2 Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 11 中的图像 2,2
    • Re-assemble the battery by connecting its cable to the connector (pull on cable to insure it is locked in place) then tightening the (4) 6-32 X 2 ¼ screws.

    • Re-assemble the battery compartment cover by tightening the (4) 6-32 X 5/16 screws.

  12. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 12 中的图像 1,1
    • Place the front case assembly over the vent module and tighten the (2) 4-40 X ¼ screws unto the dovetail mounting bracket. Reconnect the ribbon cable unto the PIM PCB. Make sure the two locking “ears” lock into position.

  13. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 13 中的图像 1,1
    • Attach the back case to the vent module and align cover with handle, air intake housing and dovetail mounting studs. Insert and tighten the included (2) 10- 32 Keps nuts and the (4) 6-32 X 2” screws.

  14. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 14 中的图像 1,1
    • Tighten the (4) 8-32 X 3 screws on the outer air intake.

  15. Zoll Impact Uni-Vent 731 PIM Board Kit Replacement: 步骤 15 中的图像 1,1
    • Perform HiPot Testing then perform Calibration and Functional Test using the RCS.

结论

To reassemble your device, follow these instructions in reverse order.