Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement

你所需要的

  1. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: 步骤 1 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement, Stand: 步骤 1 中的图像 2,2
    • Unplug the AC power connection cable.

  2. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 2 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 2 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 2 中的图像 3,3
    • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

    • Always grasp the hose by the connector. Do not pull on the hose itself.

    • Pull the connector away from the connector port.

  3. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 3 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 3 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 3 中的图像 3,3
    • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

    • Always grasp the cable by the connector. Do not pull on the cable itself.

    • Pull the connector away from the connector port.

  4. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 4 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 4 中的图像 2,2
    • Grasp the temperature probe and pull it up to remove it from the monitor.

  5. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 5 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 5 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 5 中的图像 3,3
    • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

  6. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 6 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 6 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 6 中的图像 3,3
    • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

  7. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 7 中的图像 1,1
    • Remove the flathead screw on the USB networking door.

    • Loosen the captive Phillips #2 screw securing the monitor to the stand.

  8. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 8 中的图像 1,1
    • Holding the monitor securely, open the USB networking door.

    • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

  9. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 9 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 9 中的图像 2,2
    • Detach any accessory USB cables from USB ports on the monitor.

  10. Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 10 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 10 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Stand Replacement: 步骤 10 中的图像 3,3
    • Slide the monitor off of the stand tray.

结论

To reassemble your device, follow these instructions in reverse order.