Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。

你所需要的

  1. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Stand: 步骤 1 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Stand: 步骤 1 中的图像 2,2
    • Unplug the AC power connection cable.

  2. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 2 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 2 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 2 中的图像 3,3
    • Place your thumb and forefinger on the blood pressure (NIBP) hose connector. Squeeze the side tabs until the connector releases.

    • Always grasp the hose by the connector. Do not pull on the hose itself.

    • Pull the connector away from the connector port.

  3. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 3 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 3 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 3 中的图像 3,3
    • Place your thumb and forefinger on the Pulse oximetry (SpO2 or combined SpO2/SpHb) cable connector. Squeeze the side tabs until the connector releases.

    • Always grasp the cable by the connector. Do not pull on the cable itself.

    • Pull the connector away from the connector port.

  4. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 4 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 4 中的图像 2,2
    • Grasp the temperature probe and pull it up to remove it from the monitor.

  5. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 5 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 5 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 5 中的图像 3,3
    • Remove the cover of the temperature module by pressing the tab and sliding the cover to the right.

  6. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 6 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 6 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 6 中的图像 3,3
    • Depress the spring tab on the temperature probe cable connector and withdraw it from the probe port.

  7. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 7 中的图像 1,1
    • Remove the flathead screw on the USB networking door.

    • Loosen the captive Phillips #2 screw securing the monitor to the stand.

  8. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 8 中的图像 1,1
    • Holding the monitor securely, open the USB networking door.

    • You may have to lift the monitor slightly to allow the door to swing out from the stand mount.

  9. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 9 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 9 中的图像 2,2
    • Detach any accessory USB cables from USB ports on the monitor.

  10. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 10 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 10 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 10 中的图像 3,3
    • Slide the monitor off of the stand tray.

  11. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Battery: 步骤 11 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Battery: 步骤 11 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Battery: 步骤 11 中的图像 3,3
    • Place the monitor upside down on a clean surface.

    • Insert a coin into the slot and push to open.

    • Choose a coin that fits comfortably into the slot.

    • Remove the battery cover.

  12. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 12 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 12 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 12 中的图像 3,3
    • Use the plastic label to remove the battery from its recess.

  13. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Rear Housing: 步骤 13 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Rear Housing: 步骤 13 中的图像 2,2
    • Open the communications door to access the lower left-hand rear housing screw.

    • Remove the four Phillips #2 screws (labeled 14 in the service manual) from the rear housing.

  14. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 14 中的图像 1,2 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 14 中的图像 2,2
    • With the handle/alert bar facing you, begin to lift the rear housing from the left side, holding the right side securely.

    • Do not attempt to separate the rear housing from the monitor, it is still attached with several cables.

  15. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 15 中的图像 1,1
    • Support the rear housing while you disconnect the cables in the following steps.

  16. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 16 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 16 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 16 中的图像 3,3
    • Disconnect the fan cable (labeled connector A in the service manual) from its socket in the power supply.

  17. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 17 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 17 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 17 中的图像 3,3
    • Unplug the small USB connector from its socket (labeled J4 in the service manual).

  18. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 18 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 18 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 18 中的图像 3,3
    • Disconnect the communications power cable from its socket (labeled J50 in the service manual).

    • Only pull on the connector itself and not the socket. This connection is very tight, work the connector slowly and gently.

  19. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 19 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 19 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 19 中的图像 3,3
    • Unplug the large USB connector from its socket (labeled J8 in the service manual).

  20. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 20 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 20 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 20 中的图像 3,3
    • Press the tab on the ethernet cable and unplug it from its socket (labeled J9 in the service manual).

  21. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 21 中的图像 1,1
    • The following connector requires two hands to disconnect. Flip the rear case over, and lay it on its back to gain access to the power supply cable.

  22. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 22 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 22 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 22 中的图像 3,3
    • The power supply cable is secured by an interlocking connector that must be held open to unplug the connector.

    • Insert the flat end of a spudger down into the power supply cover between the socket clip and the power supply connector.

  23. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 23 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 23 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 23 中的图像 3,3
    • Grasp the power supply cable firmly and lift both it and the spudger from the channel in the power supply housing to unplug the connector.

  24. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Module: 步骤 24 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Module: 步骤 24 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement, Module: 步骤 24 中的图像 3,3
    • Disconnect the USB cable from the module.

    • For the extended chassis: Disconnect the two USB cables from the module.

  25. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 25 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 25 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 25 中的图像 3,3
    • Remove the USB cable from the retaining clip on the module.

  26. Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 26 中的图像 1,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 26 中的图像 2,3 Welch Allyn Connex Vital Signs Monitor 6000 Module Replacement: 步骤 26 中的图像 3,3
    • Slide the module out of the case.

    • Repeat the above procedure to remove any additional modules from the chassis.

    • During reassembly, install the module closest to the display first.

    • Reinsert modules into the housing with the serial number label facing up.

结论

To reassemble your device, follow these instructions in reverse order.