The O2 and Air/Heliox filters are accessible from the rear panel of the ventilator. According to the service manual, the filters should be inspected and replaced as required.

PLEASE NOTE: This repair guide was developed by a student iFixit team based on Viasys/CareFusion's own service manual. Neither iFixit nor this repair guide is endorsed by or affiliated with Viasys or CareFusion.

Replacing these gas filters is discussed in service manual revision D:

  • Chapter 8: Preventative Maintenance, Pg 163-164
  1. Use the filter removal tool to unscrew the two filter covers. Use the filter removal tool to unscrew the two filter covers.
    • Use the filter removal tool to unscrew the two filter covers.

  2. Use the needle nosed pliers to grab the filter and pull straight out of the filter port.
    • Use the needle nosed pliers to grab the filter and pull straight out of the filter port.

    • Repeat for the second filter.

  3. Using the needle nosed pliers, grab a new filter and slide it into the port so it is correctly aligned.
    • Using the needle nosed pliers, grab a new filter and slide it into the port so it is correctly aligned.

    • Repeat for the second filter.

  4. Align the filter cover with the filter inside the port and screw back on.
    • Align the filter cover with the filter inside the port and screw back on.

    • Repeat for the second filter.