Viasys Avea Exhalation Corner Assembly Removal

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。
    • Check to see if the exhalation filter assembly is installed, if it is, remove the exhalation filter assembly.

    • Rotate the metal locking lever on the lower right of the ventilator body forward to an open position.

    • Remove the exhalation filter assembly from the ventilator body by pulling straight down.

    • Using a long Phillips screwdriver, remove the Phillips screw located at the top of the exhalation filter assembly well, if applicable.

    • Using a long Phillips screwdriver, remove the Phillips screw located at the exhalation port marked EXH, if applicable.

    • Using a long Phillips screwdriver, remove the Phillips screw from the outside of the plastic cover.

    • Using a long Phillips screwdriver, remove the Phillips screw on the font left wall.

    • Using a long Phillips screwdriver, remove the Phillips screw on the right side of the wall.

    • Set the plastic cover aside.

    • Using a long Phillips screwdriver, remove the 3 Phillips screws on the metal re-enforcement cover, if applicable, and set aside.

    • Remove the Exhalation Corner Assembly.

结论

To reassemble your device, follow the steps in the reverse order.