  1. SOMNOcomfort and SOMNObalance Ventilator Filter Replacement, Filter: 步骤 1 中的图像 1,1
    • Locate the Filter Compartment on the back of the ventilator.

  2. SOMNOcomfort and SOMNObalance Ventilator Filter Replacement: 步骤 2 中的图像 1,1
    • Push the compartment cover (22) up (from the bottom)

  3. SOMNOcomfort and SOMNObalance Ventilator Filter Replacement: 步骤 3 中的图像 1,1
    • Remove the coarse dust filter(68)

  4. SOMNOcomfort and SOMNObalance Ventilator Filter Replacement: 步骤 4 中的图像 1,1
    • Remove the fine dust filter(67)

    • Replace the filters and close the cover on the rear of the unit.

  5. SOMNOcomfort and SOMNObalance Ventilator Filter Replacement: 步骤 6 中的图像 1,1
    • Reset the filter change indicator by pressing and holding the on/off switch(19) until the filter change indicator goes blank.

    • If the fine filter was changed due to soiling before 250 operating hours, the hours meter must be reset to zero. Proceed as follows:

    • When switching the unit on, press and hold the on/off switch 19. The filter change indicator will come on after about three seconds before going off again after another three seconds. Release the on and off switch.

结论

To reassemble your device, follow these instructions in reverse order.