Puritan Bennett 560 Exhalation Block Replacement

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。
  1. Puritan Bennett 560 Exhalation Block Replacement, Exhalation Block: 步骤 1 中的图像 1,1
    • Locate the exhalation block on the front of the ventilator.

  2. Puritan Bennett 560 Exhalation Block Replacement: 步骤 2 中的图像 1,1
    • Tilt the machine up and unscrew the knob that holds the exhalation block in place.

    • This knob can be untwisted by hand, no tools required.

  3. Puritan Bennett 560 Exhalation Block Replacement: 步骤 3 中的图像 1,1
    • Pinch the "From Patient" fitting and slide the exhalation block away from the ventilator.

  4. Puritan Bennett 560 Exhalation Block Replacement: 步骤 4 中的图像 1,1
    • After replacing or cleaning the block, perform the exhalation flow sensor calibration.

    • See 6.7.3.3 in the Service Manual for instructions on calibration.

结论

To reassemble your device, follow these instructions in reverse order.