Philips Respironics V60 Top Cover Replacement

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。

你所需要的

  1. Philips Respironics V60 Top Cover Replacement, shut down the ventilator: 步骤 1 中的图像 1,2 Philips Respironics V60 Top Cover Replacement, shut down the ventilator: 步骤 1 中的图像 2,2
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    • Disconnect the ventilator from the AC power.

    • Unplug the ventilator

  2. Philips Respironics V60 Top Cover Replacement: 步骤 4 中的图像 1,1
    • Disconnect the internal battery from the power harness connector.

  3. Philips Respironics V60 Top Cover Replacement, Top Cover: 步骤 5 中的图像 1,1
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

  4. Philips Respironics V60 Top Cover Replacement: 步骤 6 中的图像 1,1
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

结论

To reassemble your device, follow these instructions in reverse order.