Utilize this guide to access inside the column, a step required for other guides such as replacing the power supply. For more information on opening the column, please refer to page 11-6 in the full service manual.

配件

没有特定配件。

  1. Use a grounding cable and wrist band connected to the GALILEO or other work surface  when you are working on the GALILEO.
    • Use a grounding cable and wrist band connected to the GALILEO or other work surface when you are working on the GALILEO.

    • Remove from GALILEO: External mains power supply, Air and oxygen supplies, Patient breathing circuit (for convenience only), Flow Sensor tubes (for convenience only).

  2. Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.
    • Loosen the rear panel of the column by removing the six hexagonal-drive (Allen) screws that hold it in place.

  3. Without removing any electrical connections, open the left side of the column as wide as practical. This exposes:  GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.
    • Without removing any electrical connections, open the left side of the column as wide as practical.

    • This exposes: GALILEO’s internal power supply, The battery fuse, The diagnostic LEDs on the power supply’s main circuit board, and the back up batteries.

    • Make sure no electrical connections are touching the column.

    • Make sure all green/brown grounding (earthing) cables are securely in place.

结论

To reassemble your device, follow these instructions in reverse order.