This guide will cover the removal and replacement of the analog PCB within the Newport e500. In order to perform this operation, a screwdriver and torque wrench will be required. Be sure to keep track of all cables as they will be reused on the replacement PCB. For some helpful tips, go to page 64 in the user manual.
你所需要的
-
-
Place device so that the top cover is facing you.
-
Open the top cover.
-
-
-
Pull up and remove the flexible rubber exhalation outlet boot (push fit).
-
-
-
Remove the 2 screws and washers securing the manifold/mixer assembly
-
-
-
Remove the screw and washer that secure the exhalation manifold support to the pneumatic panel.
-
-
-
Remove the cable from the connector J1 on the analog PCB.
-
To reassemble your device, follow these instructions in reverse order.
To reassemble your device, follow these instructions in reverse order.