How to Test the Manometer on a Fisher Paykel Neopuff
15分钟
中等
医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。
简介
转到步骤 1Please refer to page 7 of the device’s Technical Manual for source information.
-
-
Lift the cover off the Max Pressure Relief Valve slightly and turn out of the way.
-
即将完成!
If the F&P Neopuff™/Perivent™ Infant Resuscitator fails any of these tests, the manometer should be regarded as inaccurate and replaced with a new manometer.
结论
If the F&P Neopuff™/Perivent™ Infant Resuscitator fails any of these tests, the manometer should be regarded as inaccurate and replaced with a new manometer.