Hamilton-T1 Oxygen Cells Replacement

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。

你所需要的

  1. Hamilton-T1 Oxygen Cells Replacement, Oxygen Cells: 步骤 1 中的图像 1,1
    • Flip device around until the back panel is exposed.

  2. Hamilton-T1 Oxygen Cells Replacement: 步骤 2 中的图像 1,1
    • Remove the screw in the bottom center of the panel, and then take the panel off.

  3. Hamilton-T1 Oxygen Cells Replacement: 步骤 3 中的图像 1,1
    • Disconnect the oxygen cell by removing the tube.

    • Then twist the cell counter-clockwise and pull to remove the oxygen cell.

    • Install the new oxygen cell and put the back cover on.

    • After the cell is installed. Run the oxygen cell calibration.

    • Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.

结论

After the installation and calibration of the oxygen cell. The Hamilton-T1 Ventilator is ready to operate.