跳转到主内容

Fisher Paykel Neopuff Valve System Replacement

医疗设备
ifixit 或本维修指南均未获得医疗设备制造商的认可或与任何制造商存在附属关系。

你所需要的

  1. Fisher Paykel Neopuff Valve System Replacement, Valve System: 步骤 1 中的图像 1,1
    • Remove the back cover, fixed by four screws.

    • Disconnect the tube from the manometer.

    • Remove the manometer by unscrewing the two retaining nuts.

    • Fit the existing manometer into the new valve assembly panel.

    • Tighten the retaining nuts.

    • Reconnect the manometer tube.

    • Refit the front panel to the back cover with the four screws from step 1.

  2. Fisher Paykel Neopuff Valve System Replacement: 步骤 8 中的图像 1,1
    • Carry out the valve system performance test described above. It is recommended to record the lot number of the new valve assembly onto the Set-up and Maintenance Checklist.

结论

To reassemble your device, follow these instructions in reverse order.