CAIRE

Breeze Concentrator

SERVICE MANUAL

Preface

SERVICE MANUAL

Breeze Oxygen Concentrator Breeze OCSI Oxygen Concentrator Breeze Messenger Oxygen Concentrator

CAIRE, Inc.

3505 County Road 42 West

Burnsville, MN 55306-3803

Customer/Technical Service:

Toll Free Phone (U.S.A.): 1-800-48 CAIRE (1-800-482-2473)

Toll Free Fax (U.S.A.): 1-888-WE CAIRE (1-888-932-2473, to place an order)

Phone: 1-952-882-5179

Fax: 1-952-882-5178

This manual covers use and maintenance of the Breeze Oxygen Concentrator, the Breeze OCSI Oxygen Concentrator, or the Breeze Messenger Oxygen Concentrator. It is intended for use by experienced personnel only.

No attempt should be made to maintain this equipment until both this manual and Patient Operating Instruction booklet have been read and fully understood.

II Table of Contents

I Preface 2
II Table of Contents 3
III Preventative Maintenance Schedule 4-6
IV Troubleshooting 7-8
V Electrical System Components 9-11
VI Compressor Replacement 12
VII ATF Module Replacement Sheet 13
VIII Service Instructions 14
♦ OCSI Calibration 14
IX Spare Parts List 15-21
X Warranty Statement 22
XI Return Policy 23
opendix A Wiring Diagram N/A
opendix B Air Flow Diagram N/A

CAIRE Ref. No.: 10844493

Model: Breeze Oxygen Concentrator

Weekly (by patient)As Needed120 Days (approx.)365 Days (approx.)Indicated By Messenger Reporter
Clean Cabinet Intake Filter (Located right side of control panel.)
Replace Cabinet Intake Filter
Check O2 Concentration at Max. Prescribed Flowrate • 1-5 LPM min. 90% • 5-6 LPM min. 87% Check Alarm Function to Verify Adequate 9v Battery Voltage

When the unit has an oxygen concentration level below the minimum specification, the following maintenance items should be reviewed:

Replace the compressor intake Hepa filter.

Replace the final filter.

Replace the 9 volt battery.

If necessary, rebuild or replace the compressor.

As needed, vacuum dust and dirt from the interior of cabinet.

CAIRE Ref. No.: 10848321

Model: Breeze OCSI Oxygen Concentrator

Weekly (by patient)As Needed120 Days (approx.)365 Days (approx.)Indicated By Messenger Reporter
Clean Cabinet Intake Filter (Located right side of control panel.)
Replace Cabinet Intake Filter
Check O2 Concentration at Max. Prescribed Flowrate • 1-5 LPM min. 90% • 5-6 LPM min. 87% Check Alarm Function to Verify Adequate 9v Battery Voltage

When the unit has an oxygen concentration level below the minimum specification, the following maintenance items should be reviewed:

Replace the compressor intake Hepa filter.

Replace the final filter.

Replace the 9 volt battery.

If necessary, rebuild or replace the compressor.

As needed, vacuum dust and dirt from the interior of cabinet.

CAIRE Ref. No.: 10961833

Model: Breeze Messenger Oxygen Concentrator

Weekly (by patient)As Needed120 Days (approx.)365 Days (approx.)Indicated By Messenger Reporter
Clean Cabinet Intake Filter (Located right side of control panel.)
Replace Cabinet Intake Filter
Check O2 Concentration at Max. Prescribed Flowrate • 1-5 LPM min. 90% • 5-6 LPM min. 87% Check Alarm Function to Verify Adequate 9v Battery Voltage

When the unit has an oxygen concentration level below the minimum specification, the following maintenance items should be reviewed:

Replace the compressor intake Hepa filter.

Replace the final filter.

Replace the 9 volt battery.

If necessary, rebuild or replace the compressor.

As needed, vacuum dust and dirt from the interior of cabinet.

◆ Verify accuracy of Messenger.

IV Troubleshooting

Audible alarm activated, Concentrator is not running, circuit breaker tripped, no alarm lights.

PROBABLE CAUSEREMEDY
Power outage, tripped circuit breaker.Reset push button circuit breaker (top control panel).
Loose/disconnected motor-run or cold start capacitor wire.Repair or reconnect.
Faulty motor-run or cold start capacitor.Replace capacitor.
Seized or partially seized compressor.Replace or repair compressor.
Defective or shorted wiring.Repair.
Defective circuit breaker.Replace circuit breaker.
Faulty power switch.Replace.

Audible alarm activated, Concentrator is not running, circuit breaker has not tripped, no alarm lights.

PROBABLE CAUSEREMEDY
Power cord not plugged in.Plug in power cord.
No power at wall power outlet.Check house circuit breaker or fuse. Repair.
Faulty unit power switch.Replace.
Loose or disconnected wiring.Repair.
Line voltage is low.Unit restart is automatic upon restoration of proper voltage (115 volts $ \pm $ 11.5 volts).
Faulty or disconnected OCSI.Replace or repair.

Audible alarm activated, Concentrator is not running, circuit breaker has not tripped, red alarm is on.

PROBABLE CAUSEREMEDY
Thermal switch has shut down compressor because of high internal temperature.• Replace/repair fan. • Remove any obstructions to intake or exhaust. • Locate unit in well ventilated area. • Possible defective thermal switch.

No oxygen flow, unit running no alarm indications.

PROBABLE CAUSEREMEDY
Humidifier or patient tubing blocked or kinked.Clean or replace humidifier, repair or replace tubing.
Tubing between ATF module and oxygen outlet blocked or kinked.Remove unit cabinet and check tubing and components for blockage. Repair or replace as necessary.
Final filter blocked.Replace.
Control valve is closed.Turn valve counter clockwise to open/increase flow.

Low oxygen flow, flow adjusting valve full open.

PROBABLE CAUSEREMEDY
Leak or restriction in tubing or component following ATF module.Repair or replace as required.
Restriction in humidifier or patient tubing.Repair or replace as required.
Restriction internal to ATF module.Using portable flowmeter check maximum delivery from ATF module outlet, if unable to get 6 lpm or greater flow, replace module.
Final filter restricted.Replace.

IV Troubleshooting

Low concentration, 70% to 85% O₂, yellow alarm (OCSI Model).

Concentration above 85% O $ _{2} $, yellow alarm (OCSI Model).

PROBABLE CAUSEREMEDY
OCSI out of calibration.Recalibrate or replace.

Audible alarm activated, Concentrator is running, red alarm light is on.

Relief